Product Data Sheet
4OXV CiTiceL®
Oxygen (O2) Gas Sensor
Part Number: AAY80-390
Product Datasheet
4OXV Oxygen CiTiceL®
Document Purpose
The purpose of this document is to present the performance specification of the 4OXV oxygen sensor.
This document should be used in conjunction with the 4OXV Characterisation Note, Operating Principles (OP02)
and the Product Safety Datasheet (PSDS 4).
The data provided in this document are valid at 20°C, 50% RH and 1013 mBar for 3 months from the date
of sensor manufacture. For guidance on sensor performance outside of these limits, please refer to the
4OXV Characterisation Note.
Output signal can drift below the lower limit over time. For guidance on the safe use of the sensor, please refer to
the Operating Principles OP02.
Doc Ref : 4OXV.indd ECN I 4473 Issue 6
13th July 2016
Page 1 of 3
Web: www.citytech.com Email sales@citytech.com Call +44(0) 23 9228 8100
The Right Sensor Can Save A Life




City Technology

4OXV Datasheet Preview

4OXV Datasheet

Oxygen Sensor

No Preview Available !

Product Data Sheet
4OXV CiTiceL®
Oxygen (O2) Gas Sensor
Part Number: AAY80-390
Key Features and Benefits
• False alarm immunity
• Enhanced response time in extreme applications
• Reliably meets stated life
• Superior environmental performance
Technical Specifications
MEASUREMENT
Technology Electrochemical
Measurement Range 1-25% vol. O2
Maximum Overload 30% vol. O2
Output Signal* 0.10 ± 0.02 mA in Air
Response Time (T90)* <15 Seconds
Zero Current (Offset)* <0.6% vol. O2
Linearity Can be considered linear in
many cases.
See Operating Principles
(OP-02) for further details.
ELECTRICAL
Recommended Load Resistor 100
Product Dimensions
MECHANICAL
Casing Material ABS
Weight <16 g
Orientation Sensitivity <0.2% vol. O2 equivalent
ENVIRONMENTAL
Operating Temperature Range -20°C to +50°C (up to 3 months
continuous across RH range)
Recommended Storage Temp 0°C to 20°C
Thermal Transient <23.5% vol. O2
(Temp. plunge +22°C to -20°C)
Operating Pressure Range Atmospheric ± 20%
Pressure Coefficient <0.02% signal/mbar
Pressure Transient <200% signal change
(60 cm H2O step change)
Relative Humidity Range (at 0°C to 20°C)
Continuous 5 to 95%RH non-condensing
Short Term 0 to 99%RH non-condensing
IMPORTANT NOTES
All tolerances ±0.15 mm unless otherwise stated.
Do not remove label. Do not solder to pins.
When installing the sensor into instrumentation,
the sensor vent hole should not be blocked. The
instrument should also be adequately vented.
LIFETIME
Long Term Output Drift* <2% signal/month
Typically <5% over operating life
Expected Operating Life Minimum 24 months in air
Storage Life 6 months in original packaging
24 months from date of despatch
If the sensor vent hole is blocked or if the instrument
is not adequately vented, sensor performance will be
compromised.
For further details, refer to Operating Principles OP02.
* Specifications are valid at 20°C, 50% RH and 1013 mBar, using City Technology recommended
circuitry. Performance characteristics outline the performance of sensors supplied within the
first 3 months. Output signal can drift below the lower limit over time.
Doc Ref : 4OXV.indd ECN I 4473 Issue 6
13th July 2016
Page 2 of 3
Web: www.citytech.com Email sales@citytech.com Call +44(0) 23 9228 8100
The Right Sensor Can Save A Life


Part Number 4OXV
Description Oxygen Sensor
Maker City Technology
PDF Download

4OXV Datasheet PDF






Similar Datasheet

1 4OXV Oxygen Sensor
City Technology





Part Number Start With

0    1    2    3    4    5    6    7    8    9    A    B    C    D    E    F    G    H    I    J    K    L    M    N    O    P    Q    R    S    T    U    V    W    X    Y    Z



Site map

Webmaste! click here

Contact us

Buy Components

Privacy Policy