• Part: CLE-0922-400
  • Description: Chlorine Sensor
  • Manufacturer: Solidsense
  • Size: 142.96 KB
Download CLE-0922-400 Datasheet PDF
Solidsense
CLE-0922-400
Classic Line 4-Cl2-200 Sensor Chlorine Sensor 0-200 ppm Performance Characteristics 0BU Part Number Nominal Range 0 to 200ppm Sensitivity Baseline(20 o C) 0.18 ± 0.05 µA/ppm < ± 0.2 µA Baseline Drift (-20 to 40 o C) 0 to -0.3 ppm equivalent Resolution 0.1 ppm Response Time (T90) Linearity ≤ 30 seconds Linear Long Term Output Drift < 2% signal/month Operation Conditions 1BU Temperature Range -20 o C to 50 o C Operating Humidity 15 to 90%RH non-condensing Pressure Range 90 to 110 k Pa Bias Potential 0 m V Storage Life Storage Temperature 6 months in sealed container 0 o C to 20 o C Expected Operating Life 2 years in air Warranty 12 months from date of despatch Physical Characteristics 2BU Weight 5 g (approx) Orientation Sensitivity None Note: PCB sockets are remended for the sensor pin connection. Soldering to the sensor should be avoided. Note: All performance specifications are based upon the following environment conditions: 20 o C, 50% relative...