THM10-2412 Datasheet Preview

THM10-2412 Datasheet

DC/DC Converter

No Preview Available !

DC/DC Converter
Wide 2:1 input voltage 10 W DC/DC
THM 10 Series, 10 Watt
converter in a compact DIP-24 plastic case
I/O isolation 5000 VACrms rated for
250 VACrms working voltage
Certification according to IEC/EN/ES
60601-1 3rd edition for 2xMOPP
CB
Risk management process according to
Scheme
IEC 606010-1 ES 60601-1
ISO 14971 including risk management file
Acceptance criteria for electronic
The THM-10 series is a range of medical 10 Watt DC/DC converters in DIP-24
plastic package and with wide 2:1 input voltage range. They provide a reinforced
assemblies according to IPC-A-610 Level 3 isolation system for 5000 VACrms isolation and a very low leakage current of
Low leakage current < 2µA
Extended operating temperature range
–40°C to 90°C.
EMC compliance to IEC 60601-1-2 4th
edition and EN55032 class A
Operating up to 5000m altitude
5 year product warranty
less than 2 µA. The units are approved to IEC/EN/ES 60601-1 3rd edition for 2
x MOPP (Means Of Patient Protection) and come along with an ISO 14971 risk
management file. Design and production conform to the quality management
system ISO 13485. With a high efficiency of up to 89% and highest grade com-
ponents the converters can reliably operate in an ambient temperature range
of –40°C up to +90°C. They constitute a reliable solution not only for medical
equipment but also for demanding ranges of application such as transportation,
control & measurement or IGBT drivers.
Models
Order code*
Input voltage range
Output voltage
THM 10-0510
3.3 VDC
THM 10-0511
5.0 VDC
THM 10-0512
12 VDC
THM 10-0513
4.5 – 9 VDC
15 VDC
THM 10-0515
(5 VDC nominal)
24 VDC
THM 10-0521
±5.0 VDC
THM 10-0522
±12 VDC
THM 10-0523
±15 VDC
THM 10-1210
3.3 VDC
THM 10-1211
5.0 VDC
THM 10-1212
12 VDC
THM 10-1213
9. 0 – 18 VDC
15 VDC
THM 10-1215
(12 VDC nominal)
24 VDC
THM 10-1221
±5.0 VDC
THM 10-1222
±12 VDC
THM 10-1223
±15 VDC
THM 10-2410
3.3 VDC
THM 10-2411
5.0 VDC
THM 10-2412
12 VDC
THM 10-2413
18 – 36 VDC
15 VDC
THM 10-2415
(24 VDC nominal)
24 VDC
THM 10-2421
±5.0 VDC
THM 10-2422
±12 VDC
THM 10-2423
±15 VDC
THM 10-4810
3.3 VDC
THM 10-4811
5.0 VDC
THM 10-4812
12 VDC
THM 10-4813
36 – 75 VDC
15 VDC
THM 10-4815
(48 VDC nominal)
24 VDC
THM 10-4821
±5.0 VDC
THM 10-4822
±12 VDC
THM 10-4823
±15 VDC
* suffix -B2 for trim option with adjustable output
suffix -B3 for remote control option
suffix -B4 for trim + remote-control option
Output current max.
2500 mA
2000 mA
830 mA
670 mA
416 mA
±1000 mA
±416 mA
±333 mA
2500 mA
2000 mA
830 mA
670 mA
416 mA
±1000 mA
±416 mA
±333 mA
2500 mA
2000 mA
830 mA
670 mA
416 mA
±1000 mA
±416 mA
±333 mA
2500 mA
2000 mA
830 mA
670 mA
416 mA
±1000 mA
±416 mA
±333 mA
Efficiency typ.
80.0 %
84.0 %
86.5 %
87.0 %
85.5 %
83.0 %
85.5 %
86.5 %
83.0 %
85.5 %
88.0 %
89.0 %
89.0 %
84.0 %
89.0 %
88.0 %
83.0 %
86.5 %
89.0 %
89.0 %
89.0 %
85.0 %
89.0 %
88.0 %
82.5 %
86.5 %
89.0 %
89.0 %
88.5 %
85.0 %
88.0 %
88.0 %
www.tracopower.com
Page 1 of 5




TRACO

THM10-2412 Datasheet Preview

THM10-2412 Datasheet

DC/DC Converter

No Preview Available !

THM 10 Series, 10 Watt
Input Specifications
Input current no load
5 Vin models: 20 mA typ.
12 Vin models: 10 mA typ.
24 Vin models: 6 mA typ.
48 Vin models: 4 mA typ.
Surge voltage (3 s max.)
5 Vin models: 16 V max.
12 Vin models: 25 V max.
24 Vin models: 50 V max.
48 Vin models: 100 V max.
Start-up voltage
5 Vin models: 4.5 VDC (or lower)
12 Vin models: 9 VDC (or lower)
24 Vin models: 18 VDC (or lower)
48 Vin models: 36 VDC (or lower)
Startup time 30 ms typ.
Under voltage shut down
5 Vin models: 3 – 4.4 VDC
12 Vin models: 7 – 8.8 VDC
24 Vin models: 15 – 17.5 VDC
48 Vin models: 31 – 34.5 VDC
Conducted noise
– Conducted & Radiated input suppression EN 55011 limits to IEC 60601-1-2 4th editon
EN 55032 class A (internal filter)
EN 55032 class B (with external components)
– Filter proposal for compling to EN 55032 class B www.tracopower.com/overview/thm10
EMC immunity
– Generic for Medical equipment IEC/EN 60601-1-2 4th edition
– ESD (electrostatic discharge) EN 61000-4-2, air ±15 kV, contact ±8 kV,
perf. criteria A
– Radiated immunity EN 61000-4-3, 10 V/m, perf. criteria A
– Fast transient / surge EN 61000-4-4, ±2 kV, perf. criteria A
(with external input capacitor / diode) EN 61000-4-5, ±2 kV perf. criteria A
5 Vin models: Nippon chemi-con KY 1000 µF/ 25 V and
reverse diode (Vishay V10P45) in parallel
12 & 24 Vin models: Nippon chemi-con KY 470 µF/ 50 V
48 Vin models: Nippon chemi-con KY 330 µF/ 100 V
– Conducted immunity EN 61000-4-6, 10 Vrms, perf. criteria A
– Magnetic field immunity EN 61000-4-8
100 A/m, continuous, perf. criteria A
1000 A/m, 1 sec., perf. criteria A
External input fuse required
(recommended values, slow blow type)
5 Vin models: 5 A
12 Vin models: 2 A
24 Vin models: 1 A
48 Vin models: 0.5 A
www.tracopower.com
Page 2 of 5


Part Number THM10-2412
Description DC/DC Converter
Maker TRACO
Total Page 5 Pages
PDF Download

THM10-2412 Datasheet PDF

View PDF for Mobile








Similar Datasheet

1 THM10-2410 DC/DC Converter
TRACO
2 THM10-2410WI DC/DC Converters
TRACO
3 THM10-2411 DC/DC Converter
TRACO
4 THM10-2411WI DC/DC Converters
TRACO
5 THM10-2412 DC/DC Converter
TRACO
6 THM10-2412WI DC/DC Converters
TRACO
7 THM10-2413 DC/DC Converter
TRACO
8 THM10-2413WI DC/DC Converters
TRACO
9 THM10-2415 DC/DC Converter
TRACO





Part Number Start With

0    1    2    3    4    5    6    7    8    9    A    B    C    D    E    F    G    H    I    J    K    L    M    N    O    P    Q    R    S    T    U    V    W    X    Y    Z

Site map

Webmaste! click here

Contact us

Buy Components

Privacy Policy