900,000+ datasheet pdf search and download

Datasheet4U offers most rated semiconductors data sheet pdf






Tektronix

MDO3000 Datasheet Preview

MDO3000 Datasheet

Mixed Domain Oscilloscopes

No Preview Available !

Mixed Domain Oscilloscopes
MDO3000 Series Datasheet
Logic Analyzer (Optional)
16 digital channels
10 M record length on all channels
121.2 ps timing resolution
Protocol Analyzer (Optional)
Serial bus support for I2C, SPI, RS-232/422/485/UART, USB 2.0,
CAN, LIN, FlexRay, MIL-STD-1553, and Audio standards
Digital Voltmeter (Free with product registration)
4-digit AC RMS, DC, and AC+DC RMS voltage measurements
5-digit frequency measurements
Today's integrated designs need an oscilloscope that is just as integrated -
such as the MDO3000 Mixed Domain Oscilloscope (MDO) Series. It is the
ultimate 6-in-1 integrated oscilloscope that includes an integrated spectrum
analyzer, arbitrary function generator, logic analyzer, protocol analyzer, and
digital voltmeter/counter. The MDO3000 is completely customizable and
fully upgradeable. Add the instruments and performance you need now - or
later.
Key performance specifications
Oscilloscope
2 and 4 analog channel models
1 GHz, 500 MHz, 350 MHz, 200 MHz, 100 MHz bandwidth models
Bandwidth is upgradable (up to 1 GHz)
Up to 5 GS/s sample rate
10 M record length on all channels
>280,000 wfm/s maximum waveform capture rate
Standard passive voltage probes with 3.9 pF capacitive loading
and 1 GHz, 500 MHz, or 250 MHz analog bandwidth
Spectrum Analyzer
Frequency range
Standard: 9 kHz - oscilloscope bandwidth
Optional: 9 kHz - 3 GHz
Ultra-wide capture bandwidth up to 3 GHz
Arbitrary Function Generator (Optional)
13 predefined waveform types
50 MHz waveform generation
128 k arbitrary generator record length
250 MS/s arbitrary generator sample rate
Key features
FastAcqhigh-speed waveform capture rate finds elusive signal
anomalies quickly
Wave Inspector® Controls provide easy navigation and automated
search of waveform data
33 automated measurements and waveform histograms for simplified
waveform analysis
TekVPI® probe interface directly supports active, differential, and
current probes for automatic scaling and units
9 in. (229 mm) WVGA widescreen color display
Small footprint and lightweight – Only 5.8 in. (147 mm) deep and 9.2 lb.
(4.2 kg)
Spectral Analysis
Dedicated front-panel controls for commonly performed tasks
Automated peak markers identify frequency and amplitude of
spectrum peaks
Manual markers enable non-peak measurements
Trace types include: Normal, Average, Max Hold, and Min Hold
Spectrogram display enables easy observation and insight into
slowly changing RF phenomena
Automated measurements include: Channel Power, Adjacent
Channel Power Ratio (ACPR), and Occupied Bandwidth (OBW)
Arbitrary Function Generation
Generate predefined signals to quickly simulate missing devices in
designs
Capture signals on analog or digital inputs, transfer to the arbitrary
edit memory, and replicate out from the AFG
Add noise to any signal to easily perform margin testing
www.tektronix.com 1




Tektronix

MDO3000 Datasheet Preview

MDO3000 Datasheet

Mixed Domain Oscilloscopes

No Preview Available !

Datasheet
Mixed signal design and analysis
Automated triggering, decode, and search on parallel buses
Multichannel setup and hold triggering
MagniVuhigh-speed acquisition provides 121.2 ps fine timing
resolution on digital channels
Protocol Analysis
Trigger, decode and automatically search on packet-level content
on most common embedded design serial bus standards.
Export protocol decode tables for use in documenting results
Digital Voltmeter and Frequency Counter
Quickly validate voltage or frequency measurements at a glance
Graphical readout provides information on stability of the
measurement
Fully upgradeable
Add functionality, increase bandwidth or spectrum analyzer
frequency range over time as your needs change or budget allows
Optional application support
Power analysis
Limit and mask testing
Need more performance?
Need more input frequency range on the spectrum analyzer?
Need to analyze analog, digital, and RF simultaneously?
Need more record length or a larger display?
Consider the MDO4000B Series oscilloscopes www.tektronix.com/MDO4000
Oscilloscope
At the core of the MDO3000 Series is a world-class oscilloscope, offering
comprehensive tools that speed each stage of debug – from quickly
discovering anomalies and capturing them, to searching your waveform
record for events of interest and analyzing their characteristics and your
device’s behavior.
Digital phosphor technology with FastAcqhigh-
speed waveform capture
To debug a design problem, first you must know it exists. Every design
engineer spends time looking for problems in their design, a time-
consuming and frustrating task without the right debug tools.
Digital phosphor technology provides you with fast insight into the real
operation of your device. Its fast waveform capture rate – greater than
280,000 wfm/s with FastAcq – gives you a high probability of quickly seeing
the infrequent problems common in digital systems: runt pulses, glitches,
timing issues, and more.
To further enhance the visibility of rarely occurring events, intensity grading
is used to indicate how often rare transients are occurring relative to normal
signal characteristics. There are four waveform palettes available in
FastAcq acquisition mode.
The Temperature palette uses color-grading to indicate frequency of
occurrence with hot colors like red/yellow indicating frequently
occurring events and colder colors like blue/green indicating rarely
occurring events.
The Spectral palette uses color-grading to indicate frequency of
occurrence with colder colors like blue indicating frequently occurring
events and hot colors like red indicating rarely occurring events.
The Normal palette uses the default channel color (like yellow for
channel one) along with gray-scale to indicate frequency of occurrence
where frequently occurring events are bright.
The Inverted palette uses the default channel color along with gray-
scale to indicate frequency of occurrence where rarely occurring events
are bright.
These color palettes quickly highlight the events that over time occur more
often or, in the case of infrequent anomalies, occur less often.
Infinite or variable persistence choices determine how long waveforms stay
on the display, helping you to determine how often an anomaly is occurring.
► 3 GHz and 6 GHz integrated spectrum analyzers
► Simultaneous acquisition of analog, digital and RF
► 20 M record length
► 10.4 inch XGA display
2 www.tektronix.com


Part Number MDO3000
Description Mixed Domain Oscilloscopes
Maker Tektronix
PDF Download

MDO3000 Datasheet PDF






Similar Datasheet

1 MDO3000 Mixed Domain Oscilloscopes
Tektronix





Part Number Start With

0    1    2    3    4    5    6    7    8    9    A    B    C    D    E    F    G    H    I    J    K    L    M    N    O    P    Q    R    S    T    U    V    W    X    Y    Z



Site map

Webmaste! click here

Contact us

Buy Components

Privacy Policy