the N4373C for lowest noise and ripple, the accuracy has been improved by better than a factor of 2, compared to its predecessor, the 86030A 50GHz LCA. This increased accuracy improves the yield from tests performed with the N4373C by narrowing margins needed to pass the tested devices. This advanced design together with temperature-stabilized transmitter and receiver ensures repeatable measurements over hours without recalibration Relative frequency response uncertainty: ± 0.8 dB @ 50 GHz (typ) ± 1.3 dB @ 65 GHz (typ) Absolute frequency response uncertainty: ± 1.2 dB @ 50 GHz (typ) ± 1.8 dB @ 65 GHz (typ) Noise floor: -60(55) dB(A/W) for O/E measurements @ 50(65) GHz -64(59) dB(W/A) for E/O measurements @ 50(65) GHz Typical phase uncertainty: ±2.3° Transmitter wavelength: 1550nm ± 20 nm 1310nm ± 10 nm 1290 - 1610 nm with external source input Built-in optical power meter For...