Datasheet4U.com

4OXV Datasheet PDF

City Technology 4OXV

the 4OXV Characterisation Note, Operating Principles (OP02) and the Product Safety Datasheet (PSDS 4). The data provided in this document are valid at 20°C, 50% RH and 1013 mBar for 3 months from the date of sensor manufacture. For guidance on sensor performance outside of these limits, please refer to the 4OXV Characterisation Note.
Part No. 4OXV
Datasheet 4OXV Download
Manufacturer City Technology
Description Oxygen Sensor
Buy Stock & Price
4OXV datasheet page 1 preview

FEATURES AND BENEFITS

● False alarm immunity
● Enhanced response time in extreme applications
● Reliably meets stated life
● Superior environmental performance Technical Specifications MEASUREMENT Technology Electrochemical Measurement Range 1-25% vol. O2 Maximum Overload 30% vol. O2 Output Signal* 0.10 ± 0.02 mA in Air Response Time (T90)* <15 Seconds Zero Current (Offset)* <0.6% vol. O2 Linearity Can be considered linear in many cases. See Operating Principles (OP-02) for further details. ELECTRICAL Recommended Load Resistor 100 Ω Product Dimensions MECHANICAL Casing Material ABS Weight <16 g Orientation Sensitivity <0.2% vol. O2 equivalent ENVIRONMENTAL Operating Temperature Range -20°C to +50°C (up to 3 months continuous across RH range) Recommended Storage Temp 0°C to 20°C Thermal Transient <23.5% vol. O2 (Temp. plunge +22°C to -20°C) Operating Pressure Range Atmospheric ± 20% Pressure Coefficient <0.02% signal/mbar Pressure Transient <200% signal change (60 cm H2O step change) Re...
4OXV page 2 preview 4OXV page 3 preview