City Technology CNLH

the CNLH oxygen gas sensor. This document should be used in conjunction with the Operating Principles (OP02) and the Product Safety Datasheet (PSDS 4). The data provided in this document are valid at 20°C, 50% RH and 1013 mBar for 3 months from the date of sensor manufacture.
Part No. CNLH
Datasheet CNLH Download
Manufacturer City Technology
Description ppm Oxygen Sensor
Buy Stock & Price
CNLH datasheet page 1 preview
Product Data Sheet CNLH CiTiceL® Oxygen (O2) Gas Sensor Part Numbers: AA015-000 Product Datasheet CNLH ppm Oxygen Sensor (Low Hydrogen Response) Document Purpose The purpose of this document is to present the performance specification of the CNLH oxygen gas sensor. Output signal can drift below the lower limit over time. For guidance on the safe use of the sensor, please refer to the Operating Principles (OP02). Doc. Ref.: cnlh.indd ECN I 4898 Issue 3 28th August 2017 Page 1 of 3 Web: Email: sales@citytech.com Call: +44(0) 23 9228 8100 The Right Sensor Can Save A Life Product Data Sheet CNLH CiTiceL® Oxygen (O2) Gas Sensor Part Numbers: AA015-000 Performance Characteristics MEASUREMENT Operating Principle 2-electrode electrochemical Measurement Range 0-2 ppm O2 Maximum Overload 1000 ppm O2 Filter None Output Signal* 13 - 17 mV in air with a 10Ω load resistor Response Time (T95)* < 20 seconds Purge Time: < 6 hours (Ambient air to <10 ppm) Linearity Linear ELECTRICAL Recommended Load Res...