ADS1291, ADS1292, ADS1292R
SBAS502C – DECEMBER 2011 – REVISED APRIL 2020
ADS129x Low-Power, 2-Channel, 24-Bit Analog Front-End for Biopotential Measurements
1 Features
•1 Two low-noise PGAs and two high-resolution ADCs (ADS1292 and ADS1292R)
• Low power: 335 μW/channel • Input-referred noise: 8 μVPP
(150-Hz BW, G = 6) • Input bias current: 200 pA • Data rate: 125 SPS to 8 kSPS • CMRR: 120 dB • Programmable gain: 1, 2, 3, 4, 6, 8, or 12 • Supplies: unipolar or bipolar
– Analog: 2.7 V to 5.25 V – Digital: 1.7 V to 3.