Datasheet4U Logo Datasheet4U.com

ADS1293 - 24-Bit Analog Front-End

General Description

The ADS1293 incorporates all

Key Features

  • 1 Three High-Resolution Digital ECG Channels With Simultaneous Pace Output.
  • EMI-Hardened Inputs.
  • Low Power: 0.3 mW/channel.
  • Input-Referred Noise: 7 µVpp (40-Hz Bandwidth).
  • Input Bias Current: 175 pA.
  • Data Rate: Up to 25.6 ksps.
  • Differential Input Voltage Range: ±400 mV.
  • Analog Supply Voltage: 2.7 V to 5.5 V.
  • Digital I/O Supply Voltage: 1.65 V to 3.6 V.
  • Right-Leg Drive Amplifier.
  • AC and DC Lead-Off.

📥 Download Datasheet

Full PDF Text Transcription (Reference)

The following content is an automatically extracted verbatim text from the original manufacturer datasheet and is provided for reference purposes only.

View original datasheet text
Product Folder Sample & Buy Technical Documents Tools & Software Support & Community ADS1293 SNAS602C – FEBRUARY 2013 – REVISED DECEMBER 2014 ADS1293 Low-Power, 3-Channel, 24-Bit Analog Front-End for Biopotential Measurements 1 Features •1 Three High-Resolution Digital ECG Channels With Simultaneous Pace Output • EMI-Hardened Inputs • Low Power: 0.3 mW/channel • Input-Referred Noise: 7 µVpp (40-Hz Bandwidth) • Input Bias Current: 175 pA • Data Rate: Up to 25.6 ksps • Differential Input Voltage Range: ±400 mV • Analog Supply Voltage: 2.7 V to 5.5 V • Digital I/O Supply Voltage: 1.65 V to 3.