Datasheet4U Logo Datasheet4U.com

AFE5808A - 8-Channel Ultrasound Analog Front-End

General Description

The AFE5808A is a highly-integrated, analog frontend (AFE) solution specifically designed for ultrasound systems in which high performance and small size are required.

The AFE5808A integrates a complete time-gain-control (TGC) imaging path and a continuous wave Doppler (CWD) path.

Key Features

  • 8-Channel Complete Analog Front-End.
  • LNA, VCAT, PGA, LPF, ADC, and CW Mixer.
  • Programmable Gain Low-Noise Amplifier (LNA).
  • 24-, 18-, or 12-dB Gain.
  • 0.25-, 0.5-, or 1-VPP Linear Input Range.
  • 0.63-, 0.7-, or 0.9-nV/√Hz Input-Referred Noise.
  • Programmable Active Termination.
  • 40-dB, Low-Noise Voltage-Controlled Attenuator (VCAT).
  • 24-, or 30-dB Programmable Gain Amplifier (PGA).
  • 3rd-Order, Linear-Phase, Low-Pass.

📥 Download Datasheet

Full PDF Text Transcription (Reference)

The following content is an automatically extracted verbatim text from the original manufacturer datasheet and is provided for reference purposes only.

View original datasheet text
Product Folder Sample & Buy Technical Documents Tools & Software Support & Community AFE5808A SLOS729D – OCTOBER 2011 – REVISED NOVEMBER 2015 AFE5808A 0.75 nV/√Hz, 65-MSPS, 158 mW/Channel, Fully-Integrated, 8-Channel, 14- and 12-Bit, Ultrasound Analog Front-End With Passive CW Mixer 1 Features • 8-Channel Complete Analog Front-End – LNA, VCAT, PGA, LPF, ADC, and CW Mixer • Programmable Gain Low-Noise Amplifier (LNA) – 24-, 18-, or 12-dB Gain – 0.25-, 0.5-, or 1-VPP Linear Input Range – 0.63-, 0.7-, or 0.